JUMO tecLine H2O2 and PAA
The membrane-covered, amperometric sensors measure the concentration of hydrogen peroxide or peracetic acid in aqueous solutions. Typical application areas include electroplating plants, dairies, swimming pools, and the chemicals industry.
Categories: Electrodes/sensors, Liquid analysis
Technical data
| General technical data | |
|---|---|
| Measuring range | Hydrogen peroxide: max. 0.05 to 20000 mg/l (analog), max. 0.05 to 200000 mg/l (digital); peracetic acid: 0.05 to 20000 mg/l (analog and digital) |
| Measuring principle | Membrane-covered, amperometric two-electrode system with integrated electronic components |
| Inflow speed | Approx. 15 cm/s |
| Response time | Hydrogen peroxide: approx. 5 to 10 minutes; peracetic acid: approx. 5 minutes at 10 °C |
| Operating temperature | 0 to 45 °C |
| Settling time | Hydrogen peroxide: approx. 3 hrs for initial startup; peracetic acid: approx. 1 hr for initial startup |
| pH value area of application | Hydrogen peroxide: pH 2 to pH 11; peracetic acid: pH 1 to pH 6 |
| Application areas | Drinking water, swimming pool water, service water |
Downloads
BROCHURE
ORDER_INFORMATION
BROCHURE
CERTIFICATE
DECLARATION_OF_CONFORMITY
BROCHURE
TECHNICAL_BOOK
INTERFACE_DESCRIPTION
BROCHURE