JUMO tecLine CR-4P
These conductivity sensors are suitable for the measurement of electrolytic conductivity in aqueous solutions. In contrast to customary two-electrode sensors, they offer application-related advantages in the range between medium and high conductance.
Categories: Electrodes/sensors, Liquid analysis
Technical data
| General technical data | |
|---|---|
| Measuring range | 1 μS/cm to 600 mS/cm |
| Electrode material | Stainless steel (1.4435) |
| Process connection | JUMO PEKA (hygienic process connection) process connection adapter available: VARIVENT® DN 40 to 125, stainless steel 1.4435 (316L); clamp DN 25/32/40 and DN 50, stainless steel 1.4435 (316L); aseptic DN 40 and DN 50, stainless steel 1.4435 (316L); aseptic NKS DN 40, stainless steel 1.4435 (316L) |
| Process pressure | 16 bar at 25 °C; 6 bar at -10 °C and +140 °C |
| Operating temperature | -10 to +120 °C, temporarily 140 °C (sterilization) |
| Cell constant | K = 0.3 to 0.4 |
| Features | Very wide measuring range; CIP and SIP-capability; hygienic design; includes quality certificate |
| Areas of application | Rinsing processes in the food and beverage industry, pharmaceuticals and biotechnology sector; CIP and SIP applications |