JUMO AQUIS 500 CR

The product controls the electrolytic conductivity, the specific electrical resistivity, or the TDS value. Conductive two-electrode measuring cells as well as four-electrode measuring cells can be connected to the device.

Technical data

General technical data
MeasurandsConductivity, temperature, resistance, TDS value
MountingSurface or control cabinet mounting
Outputs2 analog outputs, 2 relays with changeover contact
Protection typeIP67
FeaturesMultilingual plain text operation; graphic display with backlight; P, PI, PD, and PID control functions; USP switching function according to USP<645>
Areas of applicationFor universal use